Seta
Oil Test Centrifuge
Seta
Oil Test Centrifuge is a microprocessor controlled, heated
centrifuge manufactured exclusively for Seta by Hermle
Labortechnik. The instrument is fully programmable allowing
automatic configuration to the specified oil test parameters.
A four place swing out rotor and universal bucket assembly
accommodates low cost adaptors and cushions suitable for
borosilicate glass centrifuge tubes of the types specified for
oil testing: 6 or 8 inch conical, 8 inch trace sediment or 8
inch pear-shaped.
SAFETY features
include a motor driven lid latch with a safety interlock to
prevent centrifuge operation while the lid is unlocked,
imbalance detection, and over-temperature protection for both
the motor and the chamber. The centrifuge is almost silent in
operation with measured noise levels below 65 dB (A).
Two types of trolleys are available as optional accessories to
allow easy movement of the instrument. The centrifuge meets
the requirements of EN 61010-2-020; DIN 58970 parts 1, 2 &
4; and UVV VBG 7z.
SUPPLIED WITH: rotor,
buckets and cushions.
Note: Bucket
adaptors and tubes should be selected from the method table
according to test method requirements.
ACCESSORIES
- CENTRIFUGE
REQUIRED
Choose adaptors and tubes from the following options,
dependant on the method of test
(See Table Below):
90001-2:
ADAPTOR, for 90040-0 and 90060 glassware (pack of 4)
90003-2: ADAPTOR CUSHION, for 90050-0 glassware (pack of 4)
90005-2: ADAPTOR, for 90011-0 glassware (pack of 4)
90011-0: TUBE, 100ml cone-shaped, 6 inch, graduated to 0.05ml
(pack of 6)
90040-0: TUBE, 100ml cone-shaped, 8 inch, graduated to 0.05ml
(pack of 6)
90050-0: TUBE, 100ml pear-shaped, 8 inch, tube tip graduated
to 0.01ml over the range 0 to 0.2ml (pack of 6)
90060-0: TUBE, 100ml trace-sediment, graduated to 0.005ml at
tube tip (pack of 6)
Other items
which may be required include: Funnel,
Demulsifier, Solvents,
Pipette, Stirrer, Desiccator,
OPTIONAL
90010-0:
SETAFUGE HAZARD REDUCTION SYSTEM
90014-0 LOW TROLLEY, for centrifuge, with lockable castors,
height 68cm with centrifuge
90015-0 HIGH TROLLEY, for centrifuge, with lockable castors,
height 99cm with centrifuge
90016-0 TRANSFORMER, for 110/120V 60Hz operation
90200-4: WATER BATH, ambient to 80°C, for pre-heating samples
(alternative to 90300-0)
90300-0: SETA BLOCK HEATER, for use with 90301-0 to dry heat
tubes
90301-0: BLOCK, aluminum, to hold up to four 90040-0 or
90060-0 tubes
90202-2: RACK, 6 place for 90011-0 tubes
90203-2: RACK, 6 place for 90040-0 and 90060-0 tubes
90204-2: RACK, 4 place for 90050-0 tubes
90041-2: STOPPERS (Pack of 50)
99100-2 ANALYTICAL BALANCE, for ASTM D893
Items
Recommended For Each Test Method
| SETA
No. |
D91 |
D96 |
D893 |
D1796 |
D2273 |
D2709 |
D2711 |
D4007 |
| 90040-0 |
X |
X |
X |
X |
|
|
X |
X |
| 90050-0 |
|
|
|
|
|
X |
|
|
| 90060-0 |
|
|
|
|
X |
X |
|
|
| 90011-0 |
|
X |
|
|
|
|
|
|
| 90010-0 |
X |
X |
X |
X |
X |
X |
X |
X |
| 90001-2 |
X |
X |
X |
X |
X |
X |
X |
X |
| 90003-2 |
|
|
|
|
|
X |
|
|
| 90005-2 |
|
X |
|
|
|
|
|
|
| 90041-2 |
X |
X |
X |
X |
X |
X |
X |
X |
| 90202-2 |
|
X |
|
|
|
|
|
|
| 90203-2 |
X |
X |
X |
X |
X |
X |
X |
X |
| 90204-2 |
|
|
|
|
|
X |
|
|
| 90301-0 |
X |
X |
X |
X |
X |
X |
X |
X |
| 90200-4
or 90300-0 |
X |
X |
X |
X |
X |
X |
X |
X |
| Relative
Centrifugal Force (rcf) |
|
|
|
|
|
|
|
|
| 90040-0 |
600-700
x g |
500
x g |
600-700
x g |
500-800
x g |
|
|
700
x g |
600
x g |
| 90050-0 |
|
|
|
|
|
800
x g |
|
|
| 90060-0 |
|
|
|
|
600-700
x g |
800
x g |
|
|
| 90011-0 |
|
500
x g |
|
|
|
|
|
|
| Sample
Pre-Heat Temperature |
32
to 35ēC |
60
+/-3ēC |
60
+/-5ēC |
49
+/-1ēC
or 60 +/-1ēC |
32
to 35ēC |
|
82
+/-1ēC |
60
+/-3ēC |
| Test
Temperature |
Unheated |
60
+/-3ēC |
Unheated |
60
+/-1ēC |
Unheated |
21
to 32ēC |
60
+/-1ēC |
60
+/-3ēC |
| Test
Time |
10
mins. |
5
mins. |
20
mins |
10
mins. |
10
mins. |
10
mins. |
10-15
mins. |
10
mins. |
Please
visit Lazar
Scientific for further information or please email FXG
Enterprises at sales@fxg-ent.com.


|